OMNERA®
1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 16
2020-09-14
2.4 Service and Maintenance
WARNING!
When service or maintenance is to be performed the service technician shall lock the
equipment from all energy sources.
Be aware of that there are live parts even some time after having switched off the
mains.
Due to remaining energy, always wait at least 15 seconds before working on the
System.
WARNING!
There will still be live parts even when the System is switched off.
WARNING!
The equipment must not be serviced or maintained while in use with the patient.
• Risk for personal injury.
Service and maintenance shall only be performed when no patient is present.
The equipment must be checked according to the 7
to maintain
reliability and serviceability, and to ensure the safety of the patients, the operator and third
parties.
If national rules or regulations specify more frequent checks and/or maintenance, such regu-
lations must be observed.
2.4.1 Operation, Installation and Repair
WARNING!
To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
WARNING!
Only medical-approved products shall be in the X-ray room.
Risk of electric shock to patient or user.
• No non-medical electrical devices shall be used in the x-ray room.
• Note that the monitor and the PC for the Image system, are none-medical approved
products.