OMNERA®
1000-095-071-EN Rev. 1.1
Operator´s Manual
Page 28
2020-09-14
, shows a top view of the wall stand and the zone of occupancy,
the arrow showing the direction of decreasing scatter radiation levels.
3783
Residual r
adiation area
Decreasing
Decreasing
Central beam
S = significant zone of occupancy
S
S
Fig. 2-7 Zone of occupancy
2.10.1.3 Protection against residual radiation
The remaining part of the X-ray beam after having passed the plane of the image reception
area (detector and detector holder) can be significantly high. Never stand behind the wall
stand during an exposure, see Fig. 2-7