Page 130
2020-09-14
• The button turns green to indicate that the limit is set.
Fig. 4-35 Stitching mode — upper and lower limit set
• When both limits have been defined the total length of the composite image and the
number of exposures will be shown.
• The servo mode indication light will be fixed.
10. Modify the exposure settings if necessary.
Note!
For a stitching procedure, a change of the patient size or change of exposure parameters for
the first image included in the sequence, is not kept for the following included images.
11. Activate the
Exposure
button and keep it activated during the procedure.
• The starting position is always at the top edge of the composite image. Activate the
Expo-
sure
button and keep it activated until the exposure procedure is completed.
• The System will move to the correct starting position.
• When the System is in the correct position for the first image, the first image is captured.
• After exposure, the System moves to the next, correct position and the second image will
be captured.
• This is repeated until all images for the composite image has been captured.
• The System will beep when the sequence is finished. Release the button.
4.10.4 AEC Technique Setup
For information about the AEC Technique setup, see the
Generator Manual
.