![Canon Omnera 500A Operator'S Manual Download Page 79](http://html.mh-extra.com/html/canon/omnera-500a/omnera-500a_operators-manual_61481079.webp)
Operator´s Manual
1000-095-071-EN Rev. 1.1
OMNERA®
2020-09-14
Page 71
3.2.7.2 Auto Positioning Mode
General Description
The
Auto position mode
functions as the
Free mode
with the added functionality of automatic
positioning in the room.
Automatic positions are a number of positions that can be programmed and saved into the
System. The stand will upon activation of the servo button, move to the programmed position
chosen from the imaging unit.
The mode is intended as a flexible, easy to use mode. The mode can also be used as a park-
ing mode.
The distance (H) shown in display for
Auto position mode
is the distance to the floor.
Exposure Validation
Exposure is allowed if the stand is not moving and operating properly (not in an error state).
The chosen position must have been reached successfully.
3.2.7.3 Wall Flexible Mode
General Description
The
Wall Flexible mode
is intended for examinations toward a wall stand.
The wall stand detector angle is controlled from the APR.
The OTC will upon activation of the servo button move to the programmed position associ-
ated with
Wall mode
. The stand will stop at the transport height and wait for a change in posi-
tion of the wall stand (detector height). When a change in position is detected (the user
moves/drives the wall stand up or down) the OTC will move downward and start tracking the
position of the detector.
The value is constantly updated as soon as the wall stand/OTC is moved. It is possible for an
operator to change the position so the value cannot be calculated or would be incorrect; in
those situations the display will clear the field for the value.
No Wait
At the installation of the System it is possible to select, that the OTC shall not wait for the
user to move the Wallstand before tracking starts. The OTC will then start the tracking as
soon as it reaches its final position.
Exposure Validation
Exposure is allowed if the stand is not moving, operating properly (not in an error state) and
the servo button is activated.