8 Technical specification
8.1 Electrical Characteristics
Mains voltage for the System
400 V 3N, 50/60 Hz
400 V 3~
480 V 3~
Long-time (positioning) 2 A 50/60 Hz.
Momentary (exposure):150 A, 50/60 Hz
(Short term peak value),
(recommended fuse 63 A, thermal breaker,
B curve.)
Class 1
Heat dissipation
1713 BTU/H
Generator
Series and
Mains
Voltage
Generator
Momentary
Line Current
Apparent
Mains
Resistance
Recommended Minimum
Mains
Disconnect
to Generator
(15 ft/5 m
max)
Generator
Service
Rating
Distribution
Transformer
Rating
Ground
Wire Size
50 kW
400 VAC, 3p
100 A
0.17 Ω
13.3 mm
2
100 A
65 kVa
13.3 mm
2
65 kW
400 VAC, 3p
125 A
0.13 Ω
85 kVa
80 kW
400 VAC, 3p
155 A
0.10 Ω
105 kVa
50 kW
480 VAC, 3p
80 A
0.24 Ω
65 kVa
65 kW
480 VAC, 3p
105 A
0.19 Ω
85 kVa
80 kW
480 VAC, 3p
130A
0.15 Ω
105 kVa