COPYRIGHT © CANON ELECTRONICS INC. 2008 CANON DR-X10C FIRST EDITION
3-49
CHAPTER 3 DISASSEMBLY & REASSEMBLY
4) Remove 4 cables
{
1
connected to the
ultrasonic sensor PCB.
Figure 3-709
5) Remove the registration roller.
6) Remove 3 screws
{
1
(M3, round tip) and
while unhooking 2 pairs of the fitting parts
{
2
, remove the ultrasonic sensor PCB
{
3
.
Figure 3-710
4. Pickup Unit Motor
1) Remove the pickup unit.
2) Remove 2 screws
{
1
(M3, round tip) and
remove the pickup motor
{
2
.
In the same way, remove the reverse
up/down motor
{
3
, feed motor
{
4
, and
pickup up/down motor
{
5
.
Figure 3-711