User Guide
51
"L0112HN15" (9 characters) = Reserved, defaults shown
sum
= Sum of detected and normalized backscatter, 0 ... 999. Not used in Operating System 1
CR LF
= Carriage Line Feed
This line is omitted from message 111.
LINE 5
SSSSSSSSSSSSSSSSSSSSSSSSSSSSS........(5 x 770 bytes)
CR LF
The two-way attenuated normalised backscatter profile
(100000·srad·km)
-1
unless otherwise scaled by the
Attenuated
_
SCALE
parameter. Each sample is
coded with a 20-bit HEX ASCII character set; msb nibble and bit first, 2's complement.
The profile is not corrected for the tilt angle.
CR LF
= Carriage Line Feed
This line is omitted from message 111.
LINE 6
ETX CR LF
ETX
= End-of-Text Character
CR LF
= Carriage Line Feed
NOTE
NOTE
NOTE