CS135 Ceilometer
22
Message_ID_D
Message_ID_E
Note that if low baud rates and message intervals or short polling are
required it may not be possible to output multiple longer messages.
Message_ID
is the message type to output between 0 and 999 (default
001). If
Message_ID
= 0 no message type will be output. Up to 5
messages can be set. Please refer to Section 5.1.4.
NEWLASER
No parameters
Resets the laser run time counter.
OPEN
ID
Password
Open the CS135 terminal mode.
ID
= Sensor ID as per the terminal command “ID”
Password
= The sensors user password as per the terminal command
“PASSWORD”. The default is no password.
PASSWORD
Password
Sets or clears a password from 1 to 10 characters in length. Valid ASCII
characters 0-9, a - z or A – Z and letters are case sensitive. Typing the
command PASSWORD without any parameters clears the password.
The default is no password.
POLL
Sensor_ID
Message_ID
Requests the message
Message_ID
from the sensor
Sensor_ID
.
Refer to the Section 5.1.6 for more information on this command
Note:
Message_ID
outputs the message configured by MCFG.
PSU
Blower_Voltage
Boost_Volts
Battery_Current
PSU_Present
Sets PSU battery charge parameters and external blower speed.
Blower_Voltage
sets the voltage applied to the external blower in milli-
volts (mV). From 8000 to 14000. The default is 12000.
Boost_Volts
sets the boost voltage to be applied to the PSUs battery in
mV. From 11000 to 16000. (Note: It is not recommended that the
customer alters this parameter without first consulting Campbell
Scientific.). The default is 14520.
Battery_Current
sets the charge current in milli-amps (mA). From 100
to 1500. The default is 400.
PSU_Present
sets a flag indicating if a PS135E PSU is connected to the
sensor.
PSU_Present
= 0, Power supply is not fitted.
PSU_Present
= 1, Power supply is fitted. (default)
REBOOT
No parameters
Forces a system reboot. This will restore previously saved user settings.
Any unsaved changes will be lost. (Settings are saved in terminal mode
when the command CLOSE is typed, which exits the terminal mode).
SENSITIVITY
Boundary
Detector
Alpha
Boundary
From 1 to 20, default 4
Detector
From 1 to 20, default 6
Alpha
From 0.0001 to 0.1, default 0.01
(Do not change without consulting Campbell Scientific Ltd)
SERIAL
Mode
Baud
Bits_Parity
Delay
Set or read the serial port
Mode
= 0, RS232, full duplex (default).
Mode
= 1, RS232, half duplex.
Mode
= 2, RS485, full duplex.
Mode
= 3, RS485, half duplex.
Mode
= 4, reserved.
Mode
= 5, RS422.
Baud
= 0, 300 baud.
Baud
= 1, 600 baud.
Baud
= 2, 1200 baud.
Baud
=3, 2400 baud.
Baud
= 4, 4800 baud.
Baud
= 5, 9600 baud.