User Guide
27
XXXX XXXX 1000
Photo diode calibrator has failed
XXXX XXXX 0800
The sensor could not reach the desired gain levels
XXXX XXXX 0400
Laser run time has been exceeded
XXXX XXXX 0200
Laser temperature out of range
XXXX XXXX 0100
Laser thermistor failure
XXXX XXXX 0080
Laser is obscured. This can only be set if the laser is on
XXXX XXXX 0040
Laser did not achieve significant output power
XXXX XXXX 0020
Laser max power exceeded
XXXX XXXX 0010
Laser max drive current exceeded
XXXX XXXX 0008
Laser power monitor temperature out of range
XXXX XXXX 0004
Laser power monitor test fail
XXXX XXXX 0002
Laser shutdown by top board
XXXX XXXX 0001
Laser is off
5.1.6 Message Polling
The command 'POLL
Sensor_ID Message_ID
' requests the message
Message_ID
from sensor
Sensor_ID.
Where
Message_ID
is a CS135message type as defined in Section 5.1.4.
Example of how to use the CS135 POLL command to request preconfigured message
outputs.
First configure the CS135 to use polling mode from the terminal interface via the “MCFG”
command as follows:
CS135>
MCFG 0
0 1 0 0 0 0
This configures the CS135 into polling mode leaving your message output configurations
unchanged. Exit the terminal by typing the “close” command. This will save the changes you
just made.
CS135>
CLOSE
COMMAND CLOSED
Any automatic message outputs should not have stopped. Type the “POLL” command as
shown below to verify that the system is working as expected. Note: Once you have exited
the terminal characters will not be automatically echoed back to you.
POLL 0
This should return all messages you have set via the MCFG command. In this example, only
one message output was set, the default message. So the POLL command will return text
similar to that shown below. Note: If you have no message configured (i.e. MCFG returned
0 0 0 0 0 0) then nothing will be returned from the command you just typed.
CS0100001
10 100 12345 ///// ///// ///// 000000000000