User Guide
45
Description of the most significant alarm word (left word, bits going left to right)
Bit Description
8000 XXXX XXXX
Transmitter shut-off
4000 XXXX XXXX
Transmitter failure
2000 XXXX XXXX
Receiver failure
1000 XXXX XXXX
Reserved for future use
0800 XXXX XXXX
Reserved for future use
0400 XXXX XXXX
Memory error
0200 XXXX XXXX
Light path obstruction
0100 XXXX XXXX
Receiver saturation
0080 XXXX XXXX
Reserved for future use
0040 XXXX XXXX
Reserved for future use
0020 XXXX XXXX
Reserved for future use
0010 XXXX XXXX
Reserved for future use
0008 XXXX XXXX
Reserved for future use
0004 XXXX XXXX
Reserved for future use
0002 XXXX XXXX
Reserved for future use
0001 XXXX XXXX
Ceilometer engine board failure
Description of the middle alarm word (middle word, bits going left to right)
Bit Description
XXXX 8000 XXXX
Window contamination
XXXX 4000 XXXX
Battery voltage low
XXXX 2000 XXXX
Transmitter expires
XXXX 1000 XXXX
High humidity
XXXX 0800 XXXX
Reserved for future use
XXXX 0400 XXXX
Blower failure
XXXX 0200 XXXX
Reserved for future use
XXXX 0100 XXXX
Humidity sensor failure
XXXX 0080 XXXX
Heater fault
XXXX 0040 XXXX
High background radiance
XXXX 0020 XXXX
Ceilometer engine board failure
XXXX 0010 XXXX
Reserved for future use
XXXX 0008 XXXX
Laser monitor failure
XXXX 0004 XXXX
Receiver warning
XXXX 0002 XXXX
Tilt beyond limit set by user, default 45 degrees
XXXX 0001 XXXX
Reserved for future use
Description of the least significant alarm word (right word, bits going left to right)
Bit Description
XXXX XXXX 8000
Blower is on
XXXX XXXX 4000
Blower heater is on
XXXX XXXX 2000
Internal heater is on
XXXX XXXX 1000
Working from battery
XXXX XXXX 0800
Reserved for future use
XXXX XXXX 0400
Self test in progress
XXXX XXXX 0200
Reserved for future use
XXXX XXXX 0100
Reserved for future use
XXXX XXXX 0080
Units are metres if on, else feet
XXXX XXXX 0040
Reserved for future use
XXXX XXXX 0020
Polling mode is on
XXXX XXXX 0010
Reserved for future use
XXXX XXXX 0008
Reserved for future use
XXXX XXXX 0004
Reserved for future use
XXXX XXXX 0002
Reserved for future use
XXXX XXXX 0001
Reserved for future use
Summary of Contents for CS135
Page 1: ...CS135 Ceilometer User Guide Issued 18 3 13 Copyright 2012 2013 Campbell Scientific Ltd CSL 950...
Page 2: ......
Page 4: ......
Page 6: ......
Page 19: ...User Guide 11 Fig 4 2 Setting the Tilt Angle...
Page 62: ...CS135 Ceilometer 54 LINE 3 ETX CR LF ETX End of Text Character CR LF Carriage Return Line Feed...