CS135 Ceilometer
32
0004 XXXX XXXX
Photo diode
and Laser windows are dirty. This can only be set if the laser is on
0002 XXXX XXXX
Tilt beyond limit set by user, default 45 degrees
0001 XXXX XXXX
No communications between DSP and inclinometer board
Description of the middle alarm word (middle word, bits going left to right)
Bit Description
XXXX 8000 XXXX
The sensors internal humidity is high
XXXX 4000 XXXX
Communications to the DSP board temperature and humidity chip have failed
XXXX 2000 XXXX
DSP input supply voltage is low
XXXX 1000 XXXX
Self-test active
XXXX 0800 XXXX
Watch dog counter updated
XXXX 0400 XXXX
User setting stored in flash failed their signature checks
XXXX 0200 XXXX
DSP factory calibration stored in flash has failed its signature check
XXXX 0100 XXXX
DSP board OS signature test failed
XXXX 0080 XXXX
DSP board RAM test failed
XXXX 0040 XXXX
DSP boards on board PSUs are out of bounds
XXXX 0020 XXXX
TOP board non-volatile storage is corrupt
XXXX 0010 XXXX
TOP board OS signature test has failed
XXXX 0008 XXXX
TOP boards ADC and DAC are not within specifications
XXXX 0004 XXXX
TOP boards on board PSUs are out of bounds
XXXX 0002 XXXX
Communications have failed between TOP board and the DSP
XXXX 0001 XXXX
Photo diode
background radiance is out of range
Description of the least significant alarm word (right side, bits going left to right)
Bit Description
XXXX XXXX 8000
Photo diode
temperature is out of range
XXXX XXXX 4000
Photo diode
is saturated
XXXX XXXX 2000
Photo diode
calibrator temperature is out of range
XXXX XXXX 1000
Photo diode
calibrator has failed
XXXX XXXX 0800
The sensor could not reach the desired gain levels
XXXX XXXX 0400
Laser run time has been exceeded
XXXX XXXX 0200
Laser temperature out of range
XXXX XXXX 0100
Laser thermistor failure
XXXX XXXX 0080
Laser is obscured. This can only be set if the laser is on
XXXX XXXX 0040
Laser did not achieve significant output power
XXXX XXXX 0020
Laser max power exceeded
XXXX XXXX 0010
Laser max drive current exceeded
XXXX XXXX 0008
Laser power monitor temperature out of range
XXXX XXXX 0004
Laser power monitor test fail
XXXX XXXX 0002
Laser shutdown by top board
XXXX XXXX 0001
Laser is off
CR LF
= Carriage Return and Line Feed
Summary of Contents for CS135
Page 1: ...CS135 Ceilometer User Guide Issued 18 3 13 Copyright 2012 2013 Campbell Scientific Ltd CSL 950...
Page 2: ......
Page 4: ......
Page 6: ......
Page 19: ...User Guide 11 Fig 4 2 Setting the Tilt Angle...
Page 62: ...CS135 Ceilometer 54 LINE 3 ETX CR LF ETX End of Text Character CR LF Carriage Return Line Feed...