EMC_SPEC040_CLCR.doc
Rev. E
–
2021 10 18
10/63
Install & Operating Manual
CleanSeal V3
7.2
Corrosion Resistance
Standard housing versions feature a double-layer powder coat for enhanced corrosion protection,
designed to be highly resistant to decontamination agents commonly used in Life Science
and Clean Process environments.
Camfil has chosen to evaluate corrosion resistance of its metal products with respect to
decontamination products most commonly used by our customers.
•
Hydrogen Peroxide (H
2
O
2
)
https://en.wikipedia.org/wiki/Hydrogen_peroxide
•
Peracetic Acid (CH
3
CO
2
OH)
https://en.wikipedia.org/wiki/Peracetic_acid
•
Actril
TM
Cold Sterilant
http://www.medivators.com/products-and-services/renal-systems/cold-sterilant
s
➔
A comprehensive testing report is available, upon request
➔
Corrosion resistance to other decontaminants can be determined in our laboratory, upon request
PLEASE NOTE:
CleanSeal is designed for indoor use in Life Science applications with HVAC processed air.
No corrosion resistance class according to ISO 12944 is established.
Store parts in a dry area protected from weather, prior to installation.
In environments with high condensation risk, we recommend using stainless steel versions.