Install & Operating Manual
CleanSeal V3
•
Connect a particle counter or photometer with dilution stage
to the concentration port using a D10 tube compatible with
quick-connect fittings.
Push the locking ring on the fitting to release.
•
Connect an aerosol generator to the injection port
using G1/4 fittings RIEGLER 246.32 or 246.12 or compatible.
Pull back the locking ring on the fitting to release.
Removing the Testing Shroud
•
Disconnect tubes from testing devices
Push the safety spring on the latches and
lift the latch to unlock
Make sure that the shroud safety springs remain engaged
before taking your hands off !
For ceiling housings, grip the shroud by the handles,
take the weight and compress the safety springs to
Disengage the shroud
WARNING:
Danger of injury due to falling objects
The shroud module weighs up to 10 kg (format dependent).
Improper handling can result in severe injury and property damage.
When manipulating the shroud, make sure that all safety devices properly engage
and that you follow above instructions.
•
Replace mounting frame with housing frontplate or door.
Push
Lift
Push
Disengage