Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the
user of the VisionAire should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electromagnetic
environment –
guidance
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power
supply lines
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital envi-
ronment. If the user of the VisionAire
requires continued operation during
power mains interruptions, it is recom-
mended that the VisionAire is powered
from an uninterruptible power supply
(UPS).
Power frequency
magnetic field IEC
61000-4-8
3 A / m
3 A / m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE U
T
is the A.C. mains voltage prior to application of the test level.
Summary of Contents for VisionAire 5
Page 1: ...VisionAireTM Family...
Page 246: ...VisionAire Family 246 GRE PN MN238 C4 A 21 SOC AC DC 65...
Page 248: ...VisionAire Family 248 GRE PN MN238 C4 A VisionAire VisionAire A I B lpm E H VisionAire...
Page 249: ...VisionAire Family PN MN238 C4 A 249 GRE 1 1...
Page 250: ...VisionAire Family 250 GRE PN MN238 C4 A VisionAire VISIONAIRE INTENSITY...
Page 251: ...VisionAire Family PN MN238 C4 A 251 GRE 3 1 2 I PINE SOL LESTOIL...
Page 252: ...VisionAire Family 252 GRE PN MN238 C4 A 30cm 1ft GFCI VISIONAIRE...
Page 256: ...VisionAire Family 256 GRE PN MN238 C4 A 1 2 5 3 4 5 7 6m 25ft...
Page 257: ...VisionAire Family PN MN238 C4 A 257 GRE 30cm 1ft GFCI 1 2 3 I I 5 4 5 6 10 90 5 I...
Page 258: ...VisionAire Family 258 GRE PN MN238 C4 A PINE SOL LESTOIL...
Page 259: ...VisionAire Family PN MN238 C4 A 259 GRE 30 cm 12 in...
Page 260: ...VisionAire Family 260 GRE PN MN238 C4 A VisionAire 15 30 5 cm 12 in I...