PN MN238-C4 A | User Manual ENG - 17
VisionAire Family
Alarm Conditions
All alarms are low priority alarms.
Alarm
Indicates
Action
General malfunction
yellow light
and intermittent
audible alarm
high product tank
pressure
OR
low product tank pressure
OR
high device temperature
Ensure flowmeter is open to minimum flow rate or high-
er. Ensure cannula is not kinked or obstructed. Remove
any devices connected downstream of the outlet of the
device. Ensure device has at least 12" of clearance on
all sides and intakes are not obstructed. Ensure external
gross particle intake filter is clean and not clogged. En-
sure unit is within operating temperature range. If issue
persists, contact equipment provider for service.
Oxygen monitor
yellow light
and intermittent
audible alarm
low oxygen concentration
Contact equipment provider for service.
Power failure yellow
light
and intermittent
audible alarm
power failure
Ensure device is plugged into a known, working outlet.
Ensure breaker switch is pushed in. If issue persists
contact equipment provider for service.
Summary of Contents for VisionAire 5
Page 1: ...VisionAireTM Family...
Page 246: ...VisionAire Family 246 GRE PN MN238 C4 A 21 SOC AC DC 65...
Page 248: ...VisionAire Family 248 GRE PN MN238 C4 A VisionAire VisionAire A I B lpm E H VisionAire...
Page 249: ...VisionAire Family PN MN238 C4 A 249 GRE 1 1...
Page 250: ...VisionAire Family 250 GRE PN MN238 C4 A VisionAire VISIONAIRE INTENSITY...
Page 251: ...VisionAire Family PN MN238 C4 A 251 GRE 3 1 2 I PINE SOL LESTOIL...
Page 252: ...VisionAire Family 252 GRE PN MN238 C4 A 30cm 1ft GFCI VISIONAIRE...
Page 256: ...VisionAire Family 256 GRE PN MN238 C4 A 1 2 5 3 4 5 7 6m 25ft...
Page 257: ...VisionAire Family PN MN238 C4 A 257 GRE 30cm 1ft GFCI 1 2 3 I I 5 4 5 6 10 90 5 I...
Page 258: ...VisionAire Family 258 GRE PN MN238 C4 A PINE SOL LESTOIL...
Page 259: ...VisionAire Family PN MN238 C4 A 259 GRE 30 cm 12 in...
Page 260: ...VisionAire Family 260 GRE PN MN238 C4 A VisionAire 15 30 5 cm 12 in I...