PN MN238-C4 A | User Manual ENG - 5
VisionAire Family
Safety Features
The following information will acquaint you with
safety features of the VisionAire Oxygen Concen-
trator. Make sure you read and understand all the
information contained in this manual before you
operate your unit. Should you have any questions,
your Equipment Provider will be happy to answer
them for you.
• Compressor Motor: A pressure relief valve is
fitted to the compressor outlet and is calibrated to
280 kPa (40 psig). Thermal safety is ensured by
a thermostat situated in the stator winding of the
compressor (135°C / 275 °F).
• General Malfunction: If any of the conditions
listed below occurs, the general malfunction light
(
) will illuminate and an audible intermittent
alarm will activate.
This includes:
• Obstruction to the flow of oxygen such as a
pinch or kink in the delivery cannula, trig-
gered by high product tank pressure
• High device product tank pressure condition
of greater than 33psig (±1)
• Low device product tank pressure condition
of less than 5psig (±1)
• High device temperature of greater than
135°C (275 °F), triggered by low product
tank pressure if the thermal switch located
within the compressor trips (shutting down
the compressor)
Device warning label and alarm display.
• Oxygen Monitor: The oxygen monitor detects
any drop in concentration below 82%. If this oc-
curs the low oxygen concentration warning light
(
) will illuminate. If the low O2 condition
persists, an audible intermittent alarm will also
activate in addition to the alarm light.
• Power Failure: In the event the unit is operating
and a loss of power occurs, the power warning
light (
) will illuminate and an audible
intermittent alarm will activate. See label image.
• Product Filter: ≥ 10 μm filter
Summary of Contents for VisionAire 5
Page 1: ...VisionAireTM Family...
Page 246: ...VisionAire Family 246 GRE PN MN238 C4 A 21 SOC AC DC 65...
Page 248: ...VisionAire Family 248 GRE PN MN238 C4 A VisionAire VisionAire A I B lpm E H VisionAire...
Page 249: ...VisionAire Family PN MN238 C4 A 249 GRE 1 1...
Page 250: ...VisionAire Family 250 GRE PN MN238 C4 A VisionAire VISIONAIRE INTENSITY...
Page 251: ...VisionAire Family PN MN238 C4 A 251 GRE 3 1 2 I PINE SOL LESTOIL...
Page 252: ...VisionAire Family 252 GRE PN MN238 C4 A 30cm 1ft GFCI VISIONAIRE...
Page 256: ...VisionAire Family 256 GRE PN MN238 C4 A 1 2 5 3 4 5 7 6m 25ft...
Page 257: ...VisionAire Family PN MN238 C4 A 257 GRE 30cm 1ft GFCI 1 2 3 I I 5 4 5 6 10 90 5 I...
Page 258: ...VisionAire Family 258 GRE PN MN238 C4 A PINE SOL LESTOIL...
Page 259: ...VisionAire Family PN MN238 C4 A 259 GRE 30 cm 12 in...
Page 260: ...VisionAire Family 260 GRE PN MN238 C4 A VisionAire 15 30 5 cm 12 in I...