12.5. Control
commands
12.5.1.
SRES:
Reset
Command Response
Description
_SRES_K0 _SRES_s
Reset
12.5.2.
SPAU:
Pause
Command Response
Description
_SPAU_K0 _SPAU_s
Pause
mode
12.5.3.
STBY:
Standby
Command Response
Description
_STBY_K0
_STBY_s
Standby mode for all channels
_STBY_Km
_STBY_s
Standby mode for channel m
12.5.4.
SNGA:
Open valve for zero gas calibration
Command Response
Description
_SNGA_K0
_SNGA_s
Open all three zero gas valves
_SNGA_Km
_SNGA_s
Open valve for zero gas calibration of
actual measuring range
_SNGA_Km_Mn
_SNGA_s
Open valve for zero gas calibration of
range Mn
12.5.5.
SEGA:
Open valve for end gas calibration
Command Response
Description
_SEGA_K0
_SEGA_s
Open all three end gas valves
_SEGA_Km
_SEGA_s
Open valve for end gas calibration of
actual measuring range
_SEGA_Km_Mn
_SEGA_s
Open valve for end gas calibration of
range Mn
12.5.6.
SSPL:
Purge Analyzer with zero gas
Command Response
Description
_SSPL_K0
_SSPL_s
Open zero gas valve and purge all
channels
12.5.7.
SATK:
Start automatic calibration
Command Response
Description
_SATK_Km
_SATK_s
Start automatic calibration with
selected range of channel m
_SATK_Km_Mn _SATK_s
Start automatic calibration for channel
m, Range n
12.5.8.
SEMB:
Set measuring range
Command Response
Description
_SEMB_Km_Mn
_SEMB_s
Set measuring range
Auto range will be disabled