Replace fuses with recommended fuse size indicated on rear panel of
instrument. Replacement with any other size fuse may cause damage
to the instrument and possible injury to operating personnel.
3.4.
Analog Output Connections (Appendix)
See Appendix for connector pinouts located on the analyzer rear panel. Remote range
identification and range selection are obtained via the rear panel connections. When a range
is selected, the corresponding control line is pulled low to zero VDC. Ranges not selected will
remain at approximately 5 VDC. When remote range control is selected on the front panel
switch, a contact closure is provided at the rear panel connector. Remote range selection is
made by connection of the control line for the desired range to the analyzers zero VDC line
provided in the connector. Five VDC is also provided.
3.5. Gases
1)
Nitrogen or (zero air) in pressurized cylinder.
2)
Standard span gas(es) near full-scale concentration with a nitrogen balance, in a
pressurized, certified cylinder.