Section 7
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 40 of 86
7.5. F5
Setup
From the Main Menu, F5 brings up the setup menu. Span gas concentrations, calibration
settings, range limits, alarms, password, linearization, system and measure settings can be
changed. The Setup menu begins as shown below. A description of each parameter is shown
in the information box. NOTE: Use the down arrow key to obtain the additional setup
parameters.
Figure 7-16 Setup menu screen I
Figure 7-17 setup menu screen II
7.5.1.
F1 Span Gas Concentration
Note: If you do not have a specific gas for a specific range, the calibration will use the
previous ranges calibration. You must have a least one span gas.
For calibration, it is necessary to input the concentration of the span gas in ppm. For every
range, the span gas concentration can be changed. After pressing F1 in the setup menu,
a screen appears in which changes can be made. Select with the cursor buttons the range
to change. The selected field turns black. To change parameters, switch to parameter
input mode by pressing the Enter key. The input cursor (horizontal bar under the first
character) then appears in the active edit field (black background). The cursor can be
positioned with the right and left cursor keys, and the value displayed (number or letter)
can be changed with the up and down cursor keys or entered directly. Every input has to
be concluded by pressing the Enter key again. Then the input cursor disappears and a
new range can be selected. The changes are saved by leaving the screen by pressing
"Main” or "Back.” At the right side of the screen, the range limits of the 4 ranges are
displayed. They cannot be changed in this screen.
Figure 7-18 Change Span Gas Settings