Section 3
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 22 of 86
3. Installation
3.1. General
The design of this instrument is for industrial applications. These installation instructions are
for a typical site. Direct any questions regarding specific installation situations to Technical
Service of California Analytical Instruments, Inc.
3.2.
Site and Mounting
NOTE: The carefully observe the following precautions:
Select a site free from direct sunlight, radiation from a high temperature surface, or abrupt
temperature variations.
This analyzer is not suitable for installation outdoors.
Select a site where the air is clean. Avoid exposing the instrument to corrosive or combustible
gases.
The instrument must not be subject to severe vibration. If severe vibration is present, use
isolation mounts.
The instrument is designed for rack-mounting. Optional rack mount slides are available.
Do not install near equipment emitting electromagnetic interference (EMI).
NOTE: A rear supporting brace or equivalent is required if the optional rack mount
slides were not purchased.
The power on/off switch is accessible from the rear of the instrument only. DO NOT
mount such that the power on/off switch is inaccessible.
3.3. Electrical
All wiring is connected at the rear of the instrument. The AC power is connected to the
power/fuse/switch as shown below
0
I
Figure 3-1 AC Power Switch, Connector, and Fuse.
NOTE: A defective ground may affect the operation of the instrument. The output
voltages are connected per Table 8.1.1. Shielded wiring is recommended for output
signals.