7.5.14. F6
Linearization
Pressing F6 on the Setup screen brings up the Linearization screen. The analyzer can be
linearized by a polynomial with 5 coefficients. By pressing F1, these 5 coefficients can be
changed for each range. By pressing F2, the raw value can be displayed. This is the
value before linearization and offset span correction. There are two values on the screen:
The value at the top is the linearized, offset-span-corrected value, and the other value is
the raw-value.
Figure 7-27 Linearization Screen
Figure 7-28 Coefficients Range Select
Figure 7-29 Change Coefficients
Figure 7-30 Linearized and raw data