Section 7
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 41 of 86
7.5.2. Calibration
Settings
Figure 7-19 Change Auto Calibration Settings
In the calibration settings menu, times and, deviations can be changed.
7.5.3. F1
Times
There are four times (in seconds) for auto calibration that can be changed. Purge,
measuring, calibration and verifying time. Changes are made and saved as above.
7.5.4.
F2 Measuring Deviations
During auto calibration, the measured value is only saved if it is within a certain time within
an upper and a lower limit. These two limits format a working window. In the setup menu,
the deviation is in percent.
7.5.5. F3
Deviations
Here you can change absolute and relative deviation in percent. After auto calibration, it is
checked to assure the deviations are within this limit. If the deviations are not in this limit,
a warning "Deviation error!” appears.
7.5.6.
F3 Range Limits
There are 4 different ranges. The user can define the upper range limits in ppm.
Figure 7-20 Change Range Limits