Section 12
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 80 of 86
12.4.23.
AVER:
Query Software version
Command Response
Description
_AVER_K0 _AVER_s_3MAIN_z_3USER_y_OS
MSR_x
z: Main version x.xxx.b_dd.mm.yyyy
y: User version x.xxx.b_dd.mm.yyyy
x: OSMSR version
x.xxx_dd.mm.yyyy
12.4.24.
AH2O:
QueryH
2
O correction parameter
Command Response
Description
_AH2O_Km
_AH2O_s_z.z_y.y_x.x
z.z: Dry – voltage of A in with no
water present
y.y: 1st order coefficient
x.x.: 2nd order coefficient
12.4.25.
ACO2:
Query CO
2
correction parameter
Command Response
Description
_ACO2_Km _ACO2_s_z.z_y.y_x.x_w.w
z.z: Offset – voltage of A in with no
CO
2
present
y.y: Min A in – if A in is below this
value no CO
2
correction will be done.
x.x: 1st order coefficient
w.w.: 2nd order coefficient
12.4.26.
AUPD:
Query UDP data streaming parameter
Command Response
Description
_AUDP_K0 _AUDP_s_<UDPPort>_
<DataFrequeny>_[<Mode>]
_[ <UDP_IP>]
Port: port for open the UDP
connection
DataFrequency: Frequency for
transmit the data in Hz
Mode: A: ASCII Mode
UDP_IP: Alternative IP address for
open the UDP connection when it
should not use the IP of connected
TCP/IP client
Summary of Contents for 600-NDIR