Section 12
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 78 of 86
12.4.10.
ATEM:
Temperatures
Command Response
Description
_ATEM_K0
_ATEM_s_z.z
Device temperature in degrees
Celsius is responded
_ATEM_Km _ATEM_s_z.z
Detector temperature of channel m is
returned in z.z
12.4.11.
ADRU:
Pressures/ Valve voltage
Command Response
Description
_ADRU_K0
_ADRU_s_z.z
Pressure in is responded
_ADRU_Km
_ADRU_s_z.z
EPC voltage of channel m is returned
in z.z
12.4.12.
ADUF:
Flows
Command Response
Description
_ADUF_K0 _ADUF_s_z.z_y.y_x.x
Sample gas flow of all channels is
responded
_ADUF_Km
_ADUF_s_z.z
Sample gas flow of channel m is
responded
12.4.13.
AGRD:
Polynom coefficients
Command Response
Description
_AGRD_Km_Mn _AGRD_s_Mn_a0_a1_a2_a3_a4
Polynomial coefficients of channel m
range Mn are responded
12.4.14.
AANG:
Deviation from zero point after autocalibration
Command Response
Description
_AANG_Km _AANG_s_M1_z.z_da_dr
_M2_z.z_da_dr
_M3_z.z_da_dr
_M4_z.z_da_dr
Verifying deviations from zero point
after auto calibration.
Values: measured value (z.z),
absolute dev (da), relative dev (dr)
12.4.15.
AAEG:
Deviation from end point after autocalibration
Command Response
Description
_AAEG_Km _AANG_s_M1_z.z_da_dr
_M2_z.z_da_dr
_M3_z.z_da_dr
_M4_z.z_da_dr
Verifying deviation from end point after
auto calibration
Values: measured value (z.z),
absolute dev (da), relative dev (dr)
12.4.16.
AFDA:
Purge and Autocalibration times
Command Response
Description
_AFDA_Km_SATK _AFDA_s_z_y_x_w
Auto calibration times of channel m:
z: Purge time
y: Calibration time
x: Total calibration time
w: Verify time
(z, y, x, w in seconds)
_AFDA_K0_SSPL
_AFDA_s_z
Purge time will be responded