Section 12
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 83 of 86
12.6. Settings
12.6.1.
EKAK:
The four span gas concentration values are set
Command Response
Description
_EKAK_Km_M1_w.w_M2_x.x_M3_y.y_M4_z.z _EKAK_s
Set end gas values for channel m
12.6.2.
EMBE:
The four measuring range end values are set
Command Response
Description
_EMBE_Km_ M1_w.w_M2_x.x_M3_y.y_M4_z.z
_EMBE_s
Set range limits
12.6.3.
EMBU:
The upper and the lower range switchover for auto range are set
Command Response
Description
_EMBU_Km_M1_w.w_W.W_M2_x.x_X.X_M3_y
.y_Y.Y_M4_z.z_Z.Z
_EMBU_s
Set lower and upper range
switchover limits
12.6.4.
EKEN:
Set new device identification and information
Command Response
Description
_EKEN_K0_new device-name
_EKEN_s
Set new device identification
Maximum length of device name is
40 characters
NOTE: To change device identification, you must first rename the device to “RESET”.
Now a name up to 40 characters can be given.
NOTE: The device name must not have any blanks between characters, e.g. “CAI CLD”
is not allowed. You can use underscores, e.g.. “CAI_CLD”.
12.6.5.
EGRD:
Set polynom coefficients
Command Response
Description
_EGRD_Km_Mn_a0_a1_a2_a3_a4
_EGRD_s Set polynomial coefficients of
range Mn on channel m
12.6.6.
EFDA:
Set autocalibration and purge times
Command Response
Description
_EFDA_Km_SATK_z_y_x_w
_EFDA_s
Set auto cal. times for channel m:
z: Purge time
y: Calibration time
x: Total calibration time
w: Verify time
(z, y, x, w in seconds)
_EFDA_K0_SSPL_z
_EFDA_s
Set analyzer purge time to z
seconds
12.6.7.
EPAR:
Set autocalibration tolerance values
Command Response
Description
_EPAR_Km_SATK_z.z_y.y_x.x_w.w _EPAR_s
Autocalibration tolerance value
(%):
z.z= Range 1
y.y= Range 2
x.x= Range 3
w.w= Range 4