Fault Codes
•
5. Temperature Criteria
(Optional)
If batteries are serviced under extreme temperature
conditions, temperature
sensing is recommended.
Temperature
sensing is optional and is not required
under normal temperature
conditions.
Code 52
LOW VOLTAGE (ZONE
1 DELTA TEMP)
Code 54
DELTA TEMP CHARGE
TERMINATION
Code 56
BAITERY
HOT
(ZONE 1)
Code 58
BAITERY
HOT
(ZONE 2)
Reason
Charge is terminated
in Zone 1 through Delta
Temperature.
Battery voltage is too low. Cells exhibit a
rapid rise in temperature.
Program is terminated.
What to do
Discard battery.
Reason
Charge is terminated due to rapid rise of cell
temperature
in Zone 2. Battery voltage is normal.
Program proceeds to next step. Battery is considered
fully charged.
Full-charge detection through Delta Temp and Negative
Slope coincide at about the same time when the battery
reaches full-charge.
Either method may terminate the
fast-charge.
•
What to do
Use battery as normal.
Reason
Battery temperature
has reached 45°C (113°F)
in Zone 1. Battery voltage is too low. Program is
terminated.
What to do
Discard battery.
Reason
Battery temperature
has reached 45°C (113°F)
in Zone 2. Battery voltage is normal. Program proceeds
to next step. Battery is considered fully charged.
Possible reasons:
•
Battery is old and has poor performance
•
Battery has high self discharge
•
High internal resistance (common to some ultra high
capacity NiCd cells)
What to do
Observe self-discharge.
If self-discharge
is
less than 30% in 24 hours and the capacity reading is
acceptable, the battery is usable.
5-10
•
Summary of Contents for C4000 SERIES
Page 7: ...CHAPTER 1 CHAPTER 1 INTRODUCTION 1 1 INTRODUCTION...
Page 10: ...CHAPTER 2 BASIC OPERATION CHAPTER 2 BASIC OPERATION 2 1...
Page 16: ...CHAPTER 3 PROGRAMS TARGET CAPACITY CHAPTER 3 PROGRAMS TARGET CAPACITY 3 1...
Page 26: ...CHAPTER 4 STATUS INDICATORS CHAPTER 4 STATUS INDICATORS 4 1...
Page 36: ...CHAPTER 5 CHAPTER 5 FAULT CODES 5 1 FAULT CODES...
Page 48: ...CHAPTER 6 BATTERY ADAPTERS CHAPTER 6 BATTERY ADAPTERS 6 1...
Page 53: ...CHAPTER 7 PROGRAMMING BATTERY ADAPTERS CHAPTER 7 PROGRAMMING BATTERY ADAPTERS 7 1...
Page 64: ...CHAPTERS CHAPTER 8 UTiliTIES 8 1 UTiliTIES...
Page 72: ...CHAPTER 9 CUSTOM PROGRAMS CHAPTER 9 CUSTOM PROGRAMS 9 1...
Page 84: ...CHAPTER 10 CHAPTER 10 OPTIONS 10 1 OPTIONS...
Page 94: ...CHAPTER 11 CHAPTER 11 C CODE TABLES 11 1 C CODE TABLES...
Page 116: ...CHAPTER 12 CHAPTER 12 DIAGNOSTICS 12 1 DIAGNOSTICS...
Page 119: ...CHAPTER 13 CHAPTER 13 SERVICE 13 1 SERVICE...
Page 127: ...CHAPTER 14 SPECIFICATIONS CHAPTER 14 SPECIFICATIONS 14 1...
Page 131: ...APPENDIX A 1...
Page 146: ...GLOSSARY G 1 GLOSSARY...