•
•
•
Fault Codes
Fault
Definition
Reason
Battery
What to
do
Code
Chemlstrv
42
Discharge time-out
Capacity exceeds 250%
NiCd, NiMH
Check battery rating within
capacity
C-Code
44
Charge time-out
Battery too large for allotted
NiCd, NiMH
Check battery rating within
charge time
C-Code
46
Recondition time-out
Allotted Recondition time is too
NiCd, NiMH
Check battery rating within
long due to:
C-Code
•
Cell mismatch
•
Battery protection device
•
Recondition current too low
ODtion
52
Low voltage. Charge
Shorted cells, charge terminated
NiCd, NiMH,
Discard battery
termination
by temperature
by delta temperature
rise
SLA
54
Normal voltage; Charge
Charge terminated by delta
NiCd, NiMH,
Use battery as normal
termination
by temperature
temperature
rise. Battery is
SLA
considered fully charged
56
Low voltage, hot battery
Service terminated,
battery
NiCd, NiMH,
Discard battery
reached 45°C (113°F)
SLA
58
Normal voltage, hot battery
Service terminated,
Battery
NiCd, NiMH,
Use battery as normal
reached 45°C (113°F). Battery is
SLA
considered fully charged.
60
Open fuse or bad driver
•
Battery fuse is open
• Any battery
• Replace battery fuse
•
Battery has high impedance
• NiCd, NiMH
• Repeat or discard battery
•
Battery is sulfated
• SLA
• Discard battery
62
Discharge current not met
Battery or driver cannot provide
NiCd, NiMH,
Try battery on a different
the requested discharge current
SLA
Station.
due to:
•
Wrong C-Code
• Check C.Code
•
Faulty battery
• Try different battery
•
Open fuse
• Replace battery fuse
•
Faulty driver
• Call for service
64
Charge current not met
Battery or driver cannot provide
NiCd, NiMH
Try battery on a different
the requested discharge current
Station.
due to:
•
Wrong C-Code
• Check C-Code
•
Faulty battery
• Try different battery
•
Open fuse
• Replace battery fuse
•
Faulty driver
• Call for service
92
Cell mismatch corrected
After several discharge-charge
NiCd, NiMH
Use battery as normal
cycles the cells adapted to each
other
95
Capacity improved to target
Exercise or recondition improved
NiCd, NiMH,
Use battery as normal
.
the capacity to target
SLA
Table
5.2
5-3
Summary of Contents for C4000 SERIES
Page 7: ...CHAPTER 1 CHAPTER 1 INTRODUCTION 1 1 INTRODUCTION...
Page 10: ...CHAPTER 2 BASIC OPERATION CHAPTER 2 BASIC OPERATION 2 1...
Page 16: ...CHAPTER 3 PROGRAMS TARGET CAPACITY CHAPTER 3 PROGRAMS TARGET CAPACITY 3 1...
Page 26: ...CHAPTER 4 STATUS INDICATORS CHAPTER 4 STATUS INDICATORS 4 1...
Page 36: ...CHAPTER 5 CHAPTER 5 FAULT CODES 5 1 FAULT CODES...
Page 48: ...CHAPTER 6 BATTERY ADAPTERS CHAPTER 6 BATTERY ADAPTERS 6 1...
Page 53: ...CHAPTER 7 PROGRAMMING BATTERY ADAPTERS CHAPTER 7 PROGRAMMING BATTERY ADAPTERS 7 1...
Page 64: ...CHAPTERS CHAPTER 8 UTiliTIES 8 1 UTiliTIES...
Page 72: ...CHAPTER 9 CUSTOM PROGRAMS CHAPTER 9 CUSTOM PROGRAMS 9 1...
Page 84: ...CHAPTER 10 CHAPTER 10 OPTIONS 10 1 OPTIONS...
Page 94: ...CHAPTER 11 CHAPTER 11 C CODE TABLES 11 1 C CODE TABLES...
Page 116: ...CHAPTER 12 CHAPTER 12 DIAGNOSTICS 12 1 DIAGNOSTICS...
Page 119: ...CHAPTER 13 CHAPTER 13 SERVICE 13 1 SERVICE...
Page 127: ...CHAPTER 14 SPECIFICATIONS CHAPTER 14 SPECIFICATIONS 14 1...
Page 131: ...APPENDIX A 1...
Page 146: ...GLOSSARY G 1 GLOSSARY...