Model 4000C Instruction Manual
Page 39
Setup Commands
The test parameters of the Model 4000C are originally set up to test most harnesses without any
problems. Therefore, although some of the following functions may never be needed, they are
made accessible to the user to permit the greatest versatility in specialized applications. Setup
commands are two-letter entries followed with a space and a number. The parameter is set when
ENTER is pressed. The two-letter commands and their corresponding ranges (and default
settings) are listed in the table below.
Command Description
2-letter
Command
Parameter Range
Maximum Receive Board
MB
1 to 4 (default = 4)
Maximum Transmit Point
MP
1 to 512 (default = 512)
Scan Delay
SD
1 to 9999 (default=800)
Test Clock Speed
SE
1 to 64 (default=20)
Learn Known Good Clock Speed
SL
1 to 64 (default=28)
Retest Delay
RD
1 to 9999 (default=500)
Retest Flag
RT
ON/OFF (default = ON)
Start Point
SP
1 to 512 (default = 1)
Clear Chain File
CC
Chain File
CF
[filename]
NOTE
Parameter settings are also saved with the program when it is saved.
The values remain in memory until a program with different settings is
loaded. Before a new program is learned with the Learn Know Good
function, reset of parameters to their default settings may be necessary.
Either use the RS command or run Default System Setting from the
front panel.
Maximum Receive Board - MB #
This option allows the user to change the number of receive boards scanned. Normally, the
maximum receive board is set to the number of boards in the system (four). This option is useful
when the system does not need all of the boards installed. Decreasing the maximum receive
boards scanned decreases the test time.
Range = 1 to 4
Default = System Size (4)
Maximum Transmit Point - MP #