background image

l

Put up label or warning board “Maintenance in progress” on or
next to the pump.

Start-up after maintenance

l

Remove label or warning board “Maintenance in progress”

l

Check the oil level according to the chapters “Checking the oil le-
vel”

l

Connect the pump up to the power supply

l

Make sure that the “Necessary installation instructions” are follo-
wed

l

Open the cooling liquid supply, adjust the flow rate to 12 l/min

l

Start the vacuum pump

l

Open the injection of cooling liquid in the cylinder, adjust the flow
rate of the injection of cooling liquid (INJ)

u

With the pump operating at 60 Hz/3600 rpm and an inlet
pressure < 100 mbar. Adjust the cooling liquid flow rate to
6 l/min or a quantity sufficient to ensure an exhaust gas outlet
temperature <80°C

u

Check the exhaust gas outlet temperature <80°C

l

If the vacuum pump is equipped with a solenoid valve at the inlet:

u

Open the solenoid valve

l

If the vacuum pump is equipped with a shut-off valve at the inlet:

u

Open the shut-off valve

Checking the oil

Checking the oil level

l

Make sure that the vacuum pump is switched off and cannot acci-
dentally be switched on again

l

Check the oil level on the different oil sight glasses

If the oil level is below the required level:

u

Top up with oil (see “Topping up Oil”)

If the oil level is above the required level:

u

Check the evacuation of the condensates.

u

Change the oil (see “Change the oil”)

Topping up with oil

NOTE

: Under normal conditions there should be no need to top up

with oil in-between the recommended oil change intervals. A
significant drop in the oil level indicates a malfunction (see
“Troubleshooting”).

NOTE

: The procedure of filling up with oil is identical for both sides. It

will be described in this operating instructions only for a single side.
Please repeat the following operations for the second side.

CAUTION

Fill in oil only through the oil filler holes.

CAUTION

Risk of injury (burns) with open oil filler orifice.

Risk of injury in case of badly screwed-in plugs.

Remove the oil filler plugs only when the vacuum pump is stopped.

The vacuum pump must only be operated when the oil filler plugs
are firmly tightened up and do not leak.

l

Make sure that the vacuum pump is switched off and cannot acci-
dentally be switched on again

l

Remove the oil filler plugs (OFP)

l

Top up with oil

l

Make sure that the oil level is on MAX-mark of the oil sight glass

l

Make sure that the seal of the oil filler plug are not damaged and
replace plug if necessary

l

Refit the oil filler plug and tighten up

Checking the colour of the oil

NOTE

: The original colour of the oil used on the COBRA vacuum

pumps is dark. Under operation, no indication can be provided by
means of the colour.

Life-span of the Oil

The oil life depends very much on the ambient and operating
conditions. Under normal conditions the oil must be changed every
5000 operating hours.

Under very unfavourable operating conditions the oil life can be less
than 500 operating hours. A very short oil life time indicates either a
malfunction (see “Troubleshooting”) or unsuitable operating
conditions.

If there is no experience available regarding the oil life under the
prevailing operation conditions, it is recommended to have the oil
analysed every 500 operating hours and thus to establish the oil
change interval accordingly.

Oil change

NOTE

: The procedure of oil change is identical for both sides. It will be

described in this operating instructions only for a single side. Please
repeat the following operations for the second side.

WARNING

In case the vacuum pump has conveyed gases that have been
contaminated with harmful foreign materials, the oil will also be
contaminated.

Danger to health during the change of contaminated oil.

Danger to the environment.

Wear protective equipment during the change of contaminated oil.

Contaminated oil is hazardous waste and must be disposed of se-
parately in compliance with applicable regulations.

Draining used oil

NOTE

: After switching off the vacuum pump at normal operating

temperature wait no more than 20 minutes before the oil is drained.

l

Make sure that the vacuum pump is switched off and cannot acci-
dentally be switched on again

l

Make sure that the vacuum pump is vented to atmospheric pres-
sure

l

Put a drain tray underneath the oil drain ball valve (ODV)

l

Open the oil drain ball valve (ODV)

l

Drain the oil

When the oil flow has stopped:

u

Close the oil drain ball valve (ODV)

l

Switch on the vacuum pump for a few seconds

l

Make sure that the vacuum pump is switched off and cannot acci-
dentally be switched on again

l

Open the oil drain ball valve again and drain any remaining oil

l

Close the oil drain ball valve (ODV) again

l

Dispose of the used oil in compliance with applicable regulations

Maintenance
Page 16

Summary of Contents for COBRA AC 1000 F

Page 1: ...e Instructions Screw Vacuum Pumps COBRA AC 1000 F VR Vapour Recovery ATEX version II 2G i o IIB T4 X Ateliers Busch S A Zone industrielle 2906 Chevenez Switzerland 0870561056 180809 Original instructi...

Page 2: ...First start up of the system 13 Switching off the system 13 Re start procedure following an Emergency Shut Down of the system 14 Maintenance 14 General information 14 Maintenance program 15 Weekly 15...

Page 3: ...g liquid inlet connection CLO Cooling liquid outlet connection OFP Oil filler plug OSG Oil sight glass ODV Oil drain ball valve MP Magnetic plug DV1 Ventilation G1 4 DV2 Aeration valve CLD Cooling liq...

Page 4: ...g liquid injection in the cylinder The COBRA vacuum pumps the principle of screw pumps Two parallel screws RT rotate in opposite directions in the pump body Entering gases are trapped between the pitc...

Page 5: ...l accordingly Local regulations regarding the motors and electric control elements must be observed when installing the pump in potentially explosive environments The maintenance instructions must be...

Page 6: ...careful packing Check the packaging for transport damage when the goods arrive The pump can withstand temperatures between 25 C and 55 C during transport Transportation of packaged pump Packed on a p...

Page 7: ...pe corresponds to the ambient conditions see Oil type Ambient pressure atmospheric Humidity range 20 to 95 Altitude up to 1000 m l Make sure that the cooling liquid fulfills the following require ment...

Page 8: ...scharge lines the line cross section should be larger than the exhaust flange in order to prevent a drop in the performance of the vacuum pump If you have any doubts contact your Busch representative...

Page 9: ...or PSA 0301 The sensor must be connected in such a way that the vacuum pump stops when the counter pressure reaches the value prescribed in the COBRA ATEX catalogue Technical data of the standard prod...

Page 10: ...meter FSA 0502 The contact flow meter must be connected in such a way that the vacuum pump stops when the flow drops below 2 l min It is possible to run the vacuum pump with an injection liquid flow o...

Page 11: ...talogue If the unit motor coupling hub is already fitted u Move the coupling hubs on the shafts in order to reach value E see COBRA ATEX catalogue l Lock the coupling hubs by tightening the set screws...

Page 12: ...t the cooling liquid inlet CLI and outlet CLO connections l Open the cooling liquid supply l Adjust the cooling liquid flow to 12 l min Connection of the injection liquid into the cylinder INJ NOTE Ma...

Page 13: ...e that the Installation Prerequisites are complied with and will remain so and ensure that adequate cooling is guaranteed If the pump is shut down for a longer period of time CAUTION If there is a ris...

Page 14: ...ations given below WARNING The inside of the vacuum pump must regularly be checked for de posits and be cleaned as and when necessary for example by pur ging the pump Deposits can cause friction and s...

Page 15: ...eaks have the vacuum pump repaired Busch service l Check the vacuum pump for cooling liquid leaks in case of leaks have the vacuum pump repaired Busch service In case of operation in a dusty environme...

Page 16: ...ler plugs OFP l Top up with oil l Make sure that the oil level is on MAX mark of the oil sight glass l Make sure that the seal of the oil filler plug are not damaged and replace plug if necessary l Re...

Page 17: ...ing liquid in compliance with applicable re gulations Checking the current consumption l Check the current of the motor An increased current indicates a fault see Troubleshooting Checking the motor ho...

Page 18: ...es must be disposed of separately in compliance with applicable environmental regulations When the vacuum pump comes to the end of its life it must be decontaminated CAUTION Only authorised personnel...

Page 19: ...artial clogging in the suction discharge or pressure line Remove the clogging Long suction discharge or pressure line with too small diameter Use larger diameter Internal parts worn or damaged Repair...

Page 20: ...ral weeks or months Let the vacuum pump run warm with inlet closed Oil viscosity is too high for the ambient tempe rature Use oil with a better viscosity class Improper oil quantity unsuitable oil typ...

Page 21: ...ge intervals are too long The oil was overheated Drain oil Fill in new oil see Maintenance In case the oil life is too short use oil with better heat resistance or retrofit cooling Cooling liquid flow...

Page 22: ...1 0486000766 510 O ring 2 0486511445 511 O ring 1 0486000645 512 O ring 2 0486000526 513 O ring 2 0486000636 514 O ring 2 0486000538 515 Shaft seal 1 0487553116 516 Seal 2 0486511080 732 Flat seal 1 0...

Page 23: ...ellus S2 V 68 High Performance hydraulic oil viscosity class ISO HC 68 20 60 C TOTAL AZOLLA ZS 68 viscosity class ISO VG 68 Oil quantity The quantity of oil specified in the following table is of info...

Page 24: ...l motor speed 50 Hz 60 Hz min 1 min 1 3000 3600 Noise level EN ISO 2151 50 Hz 60 Hz dB A dB A 83 85 Ambient temperature C 20 40 Ambient temperature high temperature version refer to the pump nameplate...

Page 25: ...tances to prevent hazard zones being reached by the upper and lower limbs EN 1012 1 2010 EN 1012 2 1996 A1 2009 Compressors and vacuum pumps Safety requirements Part 1 and 2 EN ISO 2151 2008 Acoustics...

Page 26: ...Note...

Page 27: ...Note...

Page 28: ...Majang myun Icheon si Gyunggi do 467 813 Republic of Korea Phone 82 31 645 2700 Fax 82 31 631 2898 e mail busch busch co kr Malaysia Busch Malaysia Sdn Bhd 4 6 Jalan Taboh 33 22 Seksyen 33 Shah Alam T...

Reviews: