44
of 95
Fi
tting
th
e loa
d
ce
ll t
o
the
m
anua
l pre
ss
4.3 Setting up the measurement chain on the ForceMaster
9110
The message "sensors have been changed" is shown on the display after a self-test.
1
Press the operating knob to clear this message.
In order to train the measurement chain it is necessary to carry out an
"autoconfig"
. As a result,
the ForceMaster 9110 automatically sets the evaluation elements for the part that is to be
pressed.
2
To carry out the autoconfig, insert the MasterCard into the card reader, and
turn the operating knob to the right until "
AutoConfig
" appears on the
display.
3
Push the operating knob in order to open the next sub-menu, "Step1 teach-
in“.
4
Press the operating knob again.
The display will show:
5
Carry out a pressing operation.
After you have pressed a part, the display will show "Step1 teach-in wait: up" before you move the
punch back to the top.
When the pressing operation is complete, the numeric display shows "Step2
>set mode“.
6
Turn the operating knob until you reach the following display.
The ForceMaster 9110 has saved the values it has learned, and is ready for measuring. You can
start production, with supervision.