Sheathing
Phenomenon
EMC basic standard
or test method
Immunity level (test + compli-
ance)
Home healthcare environment
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV¸ +/- 8 kV, +/- 15 kV;
+/- 25 kV air
High-frequency electromagnetic
fields
IEC 61000-4-3
10 V/m ; (80 MHz to 2.7 GHz; 80%
AM at 1 kHz)
High-frequency electromagnetic
fields in the immediate vicinity of
wireless communication devices
IEC 61000-4-3
See separate table zz (at the end of
this chapter)
Magnetic fields with rated power
frequencies
IEC 61000-4-8
See separate table zz (at the end of
this chapter)
AC port for supply input
Phenomenon
EMC basic standard
Immunity level (test + compli-
ance)
Home healthcare environment
Electrical fast transient distur-
bances/bursts
IEC 61000-4-4
+/- 2 kV; 100 kHz repetition frequen-
cy
Electrical surges: conductor to
conductor
IEC 61000-4-5
+/- 0,5 kV; +/- 1kV
Conducted interference induced
by high-frequency fields
IEC 61000-4-6
3 V; 0.15 MHz to 80 MHz; 6V in ISM
and amateur radio frequency bands
between 0.15 MHz and 80MHz 80%
AM at 1 kHz
Voltage dips
IEC 61000-4-11
0% UT; 1/2 period; at 0, 45, 90, 135,
180, 225, 270 and 315 degrees
0% UT; 1 period; and 70% UT; 25
periods; single-phase at 0 degrees
Celsius
Voltage interruptions
IEC 61000-4-11
0% UT; 250 periods
Product Description
Part A: General Information
21
Summary of Contents for Dali
Page 2: ...Part A General Information ...
Page 4: ......
Page 6: ...II ...
Page 30: ...Part B Operator and Technical Per sonnel ...
Page 32: ......
Page 36: ...IV ...
Page 88: ...Part C Care staff and residents ...
Page 90: ......
Page 129: ...Cleaning and Disinfection Part C Care staff and residents 37 ...
Page 130: ...Cleaning and Disinfection 38 Part C Care staff and residents ...
Page 131: ...Cleaning and Disinfection Part C Care staff and residents 39 ...