Type 8756 BATCH
Technical data
19
4.5.2
Pressure loss, MFC with on/off valve
Activation time [s]
Dosing amount [g]
Accuracy to reference [%]
Medium 100 mPas
Medium 1 mPas
Accuracy curve
Pressure [bar rel.]
Pressure [bar rel.]
Fig. 7: Dosing quantity depending on pressure and valve opening time < 100 g
4.6
Electrical data
4.6.1
Mass Flow Controller with on/off valve Type 6013
ATTENTION!
▶
For UL-certified components, only use limited power circuits of "NEC Class 2".
Operating voltage
24 V DC ±10 %
residual ripple < 2 %
Power consumption
▪
Maximum power consumption
▪
15 W
▪
Typical power consumption
▪
10 W for water and at a flow rate of 25 kg/h
Communication interface
büS and CANopen. The communication type can be chosen
with the Bürkert Communicator software.
Tab. 11: General electrical data
Minimum temperature rating of the cable
to be connected to the field wiring ter-
minals:
75 °C
Tab. 12: Temperature
4.6.2
Mass Flow Controller with on/off valve Type 6724
ATTENTION!
▶
For UL-certified components, only use limited power circuits of "NEC Class 2".