Parameter Setting menu
The Parameter Setting menu provides 3 access levels:
Viewing, Operator, and Administrator level.
Passcode protection
Access to the Calibration, Maintenance, Operator and Administrator levels is
protected by passcodes.
The passcodes can be defined or disabled at the Administrator level.
Point of measurement
Can be defined to identify the unit and can be displayed in the Diagnostics
menu.
Reference temperature
With temperature compensation activated, the measured value is calculated
to the value at the reference temperature (usually 20 oder 25 °C) using the
temperature coefficient.
Second rinsing
User-defined time during which the ”Rinsing” contact is closed at the end of
the rinsing cycle.
Secondary displays
Two small displays located below the main display in measuring mode.
The process variables to be displayed can be selected using the softkeys
underneath.
Slope
The slope of an electrode is the voltage change per pH unit.
For an ideal pH electrode, it lies at -59.2 mV/pH (25 °C).
Viewing level
Menu level of the Parameter Setting menu. Display of all device settings,
however no editing possible.
Glossary
Glossary
Summary of Contents for 8285
Page 1: ...Modular Analysis Transmitter 8285 Instruction manual...
Page 3: ...3 EC Declaration of Conformity 8285 Modular Process Analysis System...
Page 94: ......
Page 97: ...Transmetteur d analyse modulaire 8285 Mode d emploi...
Page 99: ...3 D claration de conformit europ enne Syst me de mesure et d analyse modulaire 8285...
Page 184: ...88 8285 Termes techniques Glossaire...
Page 190: ......
Page 193: ...Modularer Analysentransmitter 8285 Bedienungsanleitung...
Page 195: ...3 EG Konformit tserkl rung Modulares Analysenme system 8285...
Page 286: ......