52
8285
NAMUR signals: Current outputs
Behavior during messages: Function check, 22 mA signal
Behavior during messages
Output current
[mA]
“Fixed (22 mA)”
Function
check
22
4
“Last usable value”
Function
check
Depending on the parameter setting
(“Messages”), the current outputs switch to:
Currently measured value
Last measured value (HOLD function)
Fixed value (22 mA)
In the case of a fault a 22 mA signal can be
generated for the selected process variable
(1st primary value).
•
•
•
Function
check
“Current meas.”
Behavior during messages (Administrator)
OK
19.0 °C
7.00 pH
22mA message
Abort
Function check
Fixed 22 mA
Current meas.
Last usable value
As delivered, a “Warning” (maintenance request) message is generated
when the current range is exceeded (< 3.8 mA or > 20.5 mA).
This setting can be changed in the Parameter setting menu of the respective
measuring module at “Messages”.
To generate a “Failure” message, the limit value monitoring must be set to
“Variable limits”:
Parameter setting - <measuring module> - Messages - Variable limits -
Failure limit …
Enter the same values for the failure limits as for the current output:
Parameter setting - Module BASE - Output current - Variable Start / End.
Current outputs: Behavior during messages
Summary of Contents for 8285
Page 1: ...Modular Analysis Transmitter 8285 Instruction manual...
Page 3: ...3 EC Declaration of Conformity 8285 Modular Process Analysis System...
Page 94: ......
Page 97: ...Transmetteur d analyse modulaire 8285 Mode d emploi...
Page 99: ...3 D claration de conformit europ enne Syst me de mesure et d analyse modulaire 8285...
Page 184: ...88 8285 Termes techniques Glossaire...
Page 190: ......
Page 193: ...Modularer Analysentransmitter 8285 Bedienungsanleitung...
Page 195: ...3 EG Konformit tserkl rung Modulares Analysenme system 8285...
Page 286: ......