5.3 - Repair
M2041/5en
SBT-2400 Service Instructions
45
©BTG 2006
5.3.4.2
Change the O-ring Seal
.
5.3.4.3
Mount the Measuring Module
1. Insert the measuring module into the housing.
NOTE!
Position the measuring module so that the deflector ridge is on the same
side of the housing as the cable connection.
2. Tighten the three bolts slightly.
3. Mount the blade temporarily.
4. Align the blade with the housing.
Use a ruler on the flat area on the housing.
5. Tighten the three bolts hard.
Secure the three bolts using a weak type of locking fluid (Loctite 222).
6. Mount the circuit board. See section 5.3.3.2:
.
7. Mount the housing cover.
8. Mount the blade. See section 5.3.2.2:
.
9. Make a basic adjustment. See section 5.2.3:
Basic Force Calibration with SPC-1000
.
Spare parts required:
Measuring module complete
Tools required:
Snap ring pliers
2.5mm Allen key
4 mm Allen key (supplied in transmitter delivery)
(Small pliers)
Consumable required:
Locking paint
Silicon grease
Loctite 222
1
2