5.2 - Maintenance Routines
38
SBT-2400 Service Instructions
M2041/5en
©BTG 2006
10. Enter the noted value and press
OK (F1)
.
11. Remove the weight and turn the transmitter in the position that it will be
mounted.
12. Note the value for
Raw
.
13. Select the
air
option and press the
Edit (F3)
key.
14. Enter the noted value and press
OK (F1)
.
15. The transmitter is now ready to be installed.
NOTE!
The Raw sig Trim is a factory set value and should normally not be changed
unless measuring module have been changed.
5.2.3.2
Basic Force Calibration with SPC-1000
Prerequisites:
That the transmitter is mounted in a jaw wise with the blade
pointing sideways, and that the SPC-1000 is connected and working.
1. Press the
Configure
option.
A new window opens.
2. Press the
Shear force trim.
option.
A new window opens.
3. Wait until the
Raw signal
is stable, then press the
Set min
option.
4. Turn transmitter with the blade pointing downwards.
5. Hang the calibration weight on the blade.
6. Wait until the
Raw signal
is stable (approx. 30 minutes), then press the
Set
max
option.
7. Press the
Send page
option.
5.2.4
Trim Raw Signal using SPC-1000
1. Press the
Misc.
option.
A new window opens.
2. Press the
Factory setting
option.
A new window opens.
3. Enter the following code:
1632
Press
Enter
. A new window opens.
4. To adjust the Raw signal, press the
Trim up
or
Trim down
options. Press
Enter
.