5.2 - Maintenance Routines
36
SBT-2400 Service Instructions
M2041/5en
©BTG 2006
5.2.2
Analog Out Calibration
5.2.2.1
Analog Out Calibration with JCT-1200
1. Do a serial connection with a DMM to the output signal from the SBT-2400.
2. Press the
Menu (F1)
key, step down through the menu tree to the
Analog
output trim
window.
3. Select the
Test output
option, press
Edit (F3)
. Enter
4
mA and press
OK
(F1)
.
4. Press
Test (F2)
.
Note the value displayed on the DMM.
The value is displayed for 30 seconds after
Test (F2)
is pressed.
5. Select the
4 mA trim
option, press
Edit (F3)
. Enter the necessary trim ad-
justment to trim the display value to exactly 4 mA. Press
OK (F1)
.
6. Press
Test (F2)
again.
Note the value displayed on the DMM.
Repeat step 4 and step 5 until the displayed value is exactly 4 mA.
7. Select the
Test output
option, press
Edit (F3)
. Enter
20
mA and press
OK
(F1)
.
8. Press
Test (F2)
.
Note the value displayed on the DMM.
9. Select the
20 mA trim
option, press
Edit (F3)
. Enter the necessary trim ad-
justment to trim the display value to exactly 20 mA. Press
OK (F1)
.
10. Press
Test (F2)
again.
Note the value displayed on the DMM.
Repeat step 8 and step 9 until the displayed value is exactly 20 mA.
5.2.2.2
Analog Out Calibration using SPC-1000
1. Press the
Misc.
option.
A new window opens.
2. Press the
Factory setting
option.
A new window opens.
3. Enter the following code:
1632
Press
Enter
.
A new window opens.
4. Press the
4 mA
option.
A new window opens.
5. To adjust the 4mA signal, press the
4 up
or
4 down
options. Press
Enter
.
6. Press the
20 mA
option.
A new window opens.
7. To adjust the 20mA signal, press the
20 up
or
20 down
options. Press
En-
ter
.
Tools required:
Digital Multi-meter (DMM)
Tools required:
SPC-1000