82
Z31900_00_01
Operation
9.6.2
Inserting the Cassette
1.
Lift the cassette using its hand holds.
2.
Fit the cassette into the cassette guide.
3.
Push the cassette carefully into the SampleXpress. Avoid banging it into the Sam-
pleXpress unit or the magnet.
The engagement will be clearly audible.
The indicator lamp switches from blue to white.
4.
All positions are rescanned.
After the scan all touch buttons are white colored and enabled.
i
The cassette can be used without a cover. In this mode the control display may not be
accurate, because samples can be inserted or removed without program control.
9.7
The Settings Window
The
Settings
button opens a settings window containing two tabs:
•
A
User
tab for touch screen calibration and for setting the date and time;
•
A
Service (Bruker)
tab, which is password protected and can only be used by
Bruker service.
9.7.1
Touch Screen Calibration
The touch screen can be calibrated through the
Settings
button.
1.
Press the touch screen
Settings
button in the main menu.
The settings screen will appear:
Figure 9.40 Settings Screen
Summary of Contents for SampleXpress
Page 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Page 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Page 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Page 104: ...104 Z31900_00_01 Maintenance ...
Page 116: ...116 Z31900_00_01 Error Codes ...
Page 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Page 122: ...122 Z31900_00_01 Contact ...
Page 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Page 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Page 158: ......
Page 160: ...160 Z31900_00_01 Warning Signs ...
Page 164: ...164 Z31900_00_01 Figures ...
Page 166: ...166 Z31900_00_01 Tables ...
Page 168: ...168 Z31900_00_01 Glossary ...
Page 172: ...172 Z31900_00_01 Index ...
Page 173: ......