58
Z31900_00_01
Operation Overview
8.2
General Operating Guidelines
There are several general rules and procedures that should be observed while operating
the SampleXpress.
NMR Measurement
Do not operate the unit, e.g. exchange sample cassettes, during an NMR measurement
as this may disturb the NMR system. The laboratory supervisor should define proce-
dures regarding access to the unit and/or the 5 Gauss area around the magnet.
Sample Mix-up
Sample mix-up from improperly read sample labels or position markings may result
when adequate lighting is not provided. Use the operator access position for monitored
sample loading/unloading, or place the cassette on a desk before loading/unloading
samples.
The laboratory supervisor should define who should be allowed to load or unload sam-
ples from cassettes, and when. The operator should work carefully and, optimally, use
tube ID barcode collars.
ID Barcode
When the sample ID barcode is dirty it should be cleaned according to established
instruction or the sample ID barcode may be incorrectly read.
Temperature Changes
When the ambient temperature changes significantly (for example, after moving the
SampleXpress from the stock room before mounting), wait at least 1 hour until the
device has achieved room temperature before turning on the device.
Contamination of Samples
Reusing NMR glass tubes may result in contamination of a new NMR sample from the
old substance.
Summary of Contents for SampleXpress
Page 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Page 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Page 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Page 104: ...104 Z31900_00_01 Maintenance ...
Page 116: ...116 Z31900_00_01 Error Codes ...
Page 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Page 122: ...122 Z31900_00_01 Contact ...
Page 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Page 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Page 158: ......
Page 160: ...160 Z31900_00_01 Warning Signs ...
Page 164: ...164 Z31900_00_01 Figures ...
Page 166: ...166 Z31900_00_01 Tables ...
Page 168: ...168 Z31900_00_01 Glossary ...
Page 172: ...172 Z31900_00_01 Index ...
Page 173: ......