113
Z31900_00_01
Error Codes
103.310.123
103.310.124
103.310.125
103.310.126
103.310.127
103.310.128
Missing Cassette.
There is no cassette available
to execute the requested opera-
tion.
• Insert a cassette into the Sam-
pleXpress to allow full functionality
of the device.
103.310.143
103.310.144
Magnet already
occupied.
There is already a sample down
in the magnet. Inserting further
samples is prohibited.
• Check if there is really no sample
in the magnet.
• Adjust sample detection sensor
threshold (Bruker service only).
103.310.148
Tube ID changed
unexpected for posi-
tion XX. Already reg-
istered: XXX - New
Read: YYY
The Tube ID for this position
was read in the past and has
changed now to another Tube
ID unexpected.
• Check barcode collar Tube ID;
• Clean barcode reader detection
surface;
• Clean/replace barcode collar.
103.310.149
Insufficient air pres-
sure.
The pressure supply fell below
3.0 bars. Device has been
halted and all pneumatic valves
have been switched off.
• Check the main air pressure sup-
ply.
117.310.020
Emergency Stop
active.
The emergency stop button is
currently pressed.
• Be sure that the device and all of
its samples are in a secure state
before you release the emergency
stop button and restart the device.
117.310.022
117.310.023
117.310.024
117.310.025
117.310.113
117.310.115
117.310.116
117.310.117
Moving to destination
position failed.
The requested destination posi-
tion could not be reached.
• Check if something (e.g. broken
glass tubes) blocks the sample
conveyor chain motion.
117.310.028
117.310.030
117.310.121
Sample hovering
failed.
The sample could not be
detected to be hovering at the
lift air flow.
• Check the sensor operation and
clean the sensor detection sur-
face.
117.310.029
Open Guide failed.
The movable half shell of the
guide did not reach its open
position.
• Check guide operation manually.
117.310.031
117.410.065
117.410.143
117.410.144
Sensor Error: Sam-
ple hovering without
lift.
A sample was detected to be
hovering without the lift being
activated.
• Check the sensor operation
• Clean the sensor detection sur-
face.
117.310.032
Security Eject failed.
A sample was detected in the
magnet but could not be placed
back into the magazine.
• Check sensors working.
• Adjust sample detection sensor
threshold (Bruker service only).
Summary of Contents for SampleXpress
Page 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Page 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Page 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Page 104: ...104 Z31900_00_01 Maintenance ...
Page 116: ...116 Z31900_00_01 Error Codes ...
Page 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Page 122: ...122 Z31900_00_01 Contact ...
Page 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Page 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Page 158: ......
Page 160: ...160 Z31900_00_01 Warning Signs ...
Page 164: ...164 Z31900_00_01 Figures ...
Page 166: ...166 Z31900_00_01 Tables ...
Page 168: ...168 Z31900_00_01 Glossary ...
Page 172: ...172 Z31900_00_01 Index ...
Page 173: ......