
112
Z31900_00_01
Error Codes
103.310.107
117.310.126
Missing empty posi-
tion for security eject.
The cassette was searched for
an empty position for executing
a security eject attempt. But the
cassette is completely filled.
• Activate Sample Down Detection
Mode (Bruker service only).
This allows a completely filled cas-
sette to be added to the SampleX-
press. Otherwise at least 1 empty
position is needed in the cassette
to execute a Security Eject of a
potential sample in the magnet.
103.310.012
103.310.032
Missing empty posi-
tion.
The cassette was searched for
an empty position for executing
a security eject attempt. But the
cassette is completely filled.
• Allow at least 1 empty position in
the cassette.
• If the cassette is not filled com-
pletely, then clean the sensor sur-
faces.
• Adjust sample detection sensor
threshold (Bruker service only).
103.310.063
Sample Detect at
magnet failed.
A sample was not detected in
the magazine for the requested
eject.
Check if:
• There is really no sample in the
magnet (or the sample down
detection sensor threshold is set
up not correctly).
• The lift air flow is strong enough to
eject the sample back into the
magazine.
• The sample detection sensors
work well.
103.310.094
103.310.096
117.310.058
117.310.038
117.310.124
Missing empty posi-
tion for eject.
The position where the sample
from the magnet should be
ejected to is occupied. Allow at
least 1 empty position in the
cassette.
• Check if there is really a sample.
• Clean sensor surfaces.
• Adjust sample detection threshold
(Bruker service only).
103.310.097
Sample from mag-
net was placed in
position XX.
The source position of the sam-
ple in the magnet was
unknown.
103.310.099
103.310.100
103.310.101
Source position not
empty.
The position in the cassette
from where the sample was
inserted into the magnet is
occupied.
• Do not fill the position from the
sample down into the magnet.
103.310.103
Source position not
empty, no other
empty position
found.
The position in the cassette
from where the sample was
inserted into the magnet is
occupied and there is no other
empty position available.
• Do not fill the position from the
sample down into the magnet.
103.310.131
103.310.134
103.310.136
Missing empty posi-
tion.
There is no empty position in
the cassette for the operation
requested.
• Allow at least 1 empty position in
the cassette.
• If the cassette is not filled com-
pletely, then clean the sensor sur-
faces.
• Adjust sample detection threshold
(Bruker service only).
Summary of Contents for SampleXpress
Page 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Page 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Page 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Page 104: ...104 Z31900_00_01 Maintenance ...
Page 116: ...116 Z31900_00_01 Error Codes ...
Page 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Page 122: ...122 Z31900_00_01 Contact ...
Page 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Page 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Page 158: ......
Page 160: ...160 Z31900_00_01 Warning Signs ...
Page 164: ...164 Z31900_00_01 Figures ...
Page 166: ...166 Z31900_00_01 Tables ...
Page 168: ...168 Z31900_00_01 Glossary ...
Page 172: ...172 Z31900_00_01 Index ...
Page 173: ......