22
Z31900_00_01
Safety
3.5.6
Dangers from Gases Under Pressure
Pneumatics
Suffocation
WARNING
Danger of injury due to movements caused by stored pneumatic forces!
Pneumatically driven components may move unexpectedly due to stored residual
forces, causing serious injuries.
f
Work on the pneumatics system must only be carried out by trained pneumatics
technicians.
f
Before starting work on the pneumatics system, ensure that it has been com-
pletely depressurised. The pressure accumulator must be completely relieved.
WARNING
Accident hazard from asphyxiation!
A break in the pneumatic hose may result in the uncontrolled exit of nitrogen into the
laboratory.
f
An oxygen warning device should be present in the laboratory if the device is
operated with nitrogen.
f
Note that leakage from the main supply line cannot be stopped by the
SampleXpress!
Summary of Contents for SampleXpress
Page 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Page 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Page 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Page 104: ...104 Z31900_00_01 Maintenance ...
Page 116: ...116 Z31900_00_01 Error Codes ...
Page 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Page 122: ...122 Z31900_00_01 Contact ...
Page 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Page 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Page 158: ......
Page 160: ...160 Z31900_00_01 Warning Signs ...
Page 164: ...164 Z31900_00_01 Figures ...
Page 166: ...166 Z31900_00_01 Tables ...
Page 168: ...168 Z31900_00_01 Glossary ...
Page 172: ...172 Z31900_00_01 Index ...
Page 173: ......