93
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
Continued from page before
Indicator
Possible reason
Solution
By
The zero reading of
the helium level sen-
sor can not be adju-
sted at the beginning
of cooling down with
liquid helium.
The helium level
sensor is not con-
nected correctly with
the connector in the
helium flow system.
Check the connection in the
helium fill-in turret between
helium level sensor and con-
nector.
Bruker Service
The helium level
sensor is defective.
Check the helium level sensor
with the 0% calibration plug.
Bruker Service
The helium level
does not reach
100% after cooling
down.
The transport dewar
is empty, no more
helium is transferred.
Refill or change transport
dewar.
Bruker Service
The helium level
sensor is disturbed
by the transfer line’s
extension piece.
1.
Stop the liquid helium
transfer.
2.
Remove the transfer line.
3.
Measure the helium level
after some minutes
without the transfer line.
Bruker Service
After cool down the
helium boil off is
higher than specified
(up to 5 times).
Usual behavior.
A few days are
necessary for the
radiation shields and
the insulation to
reach scheduled
temperatures.
Wait a few days and check
helium boil off.
The presence of the current
lead in the current lead turret
during energizing and
shimming helps to cool down
the radiation shield due to
higher helium flow.
Bruker Service
Temperature of the
radiation shield
decreases too slowly
(if T
RS
> 250 K after
pre-cooling with
liquid nitrogen)
Cryogenic Refrigera-
tor not operating.
Start Cryogenic Refrigerator.
Bruker Service
Cryogenic Refrigera-
tion operating not
correct
See ”During Operation of the
Cryogenic Refrigerator” on
page 6-102.
Bruker Service
2 days after cool
down the T
RS
is still
higher than set value
(alarm of External
Temperature Control
Unit)
Cryogenic Refrigera-
tor performance not
sufficient.
See ”During Operation of the
Cryogenic Refrigerator” on
page 6-102.
Bruker Service
Alarm default set-
tings of the External
Temperature Control
Unit or the console
not correct.
Check set values (see
separate manual of the Exter-
nal Temperature Control Unit
and of the console)
Bruker Service
Cold head not moun-
ted correctly.
See chapter ”Mounting the
Cryogenic Refrigerator Parts”
on page 4-54.
Bruker Service
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...