129
ZTKS0156 / Z31820 / Rev.: 03
Disassembling
8.2.7
Assembling of the Transportation Lock
Bottom of Cryostat
Top of the Cryostat
Figure 8.7:
Assembling the transportation lock - step 1
1.
Insert brass nut (2) into the transport bottom
plate (1).
2.
Attach the bottom fixture flange (3) onto the
brass nut (2).
3.
Insert threaded bar (4) from the top.
4.
Screw threaded bar (4) into the brass nut
(2).
5.
Fix the lock nut (5).
6.
Insert the assembly of the lower transporta-
tion lock into the helium bore tube. Prevent
clamping the super-insulation with the
transportation lock.
7.
Fix the transport bottom plate (1) with twelve
M 5 x 12 screws.
8.
Insert and fix the three threaded bolts M 8
(6).
9.
Fix the three nuts M 8 (7).
Figure 8.8:
Assembling the transportation lock - step 2
10. Insert the top fixture flange (3).
11. Insert intermediate flange (2).
12. Insert short top fixture flange (1).
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...