42
ZTKS0156 / Z31820 / Rev.: 03
Assembling
Removing the lower Transportation Lock
i
Each transportation lock is adapted to only one cryostat, indicated by the magnet system
number in each part of the transportation lock.
Do not mix-up transportation locks from other cryostat. Keep all transportation lock parts
for future transportation.
For storage use the compartment of the shipping box of the cryostat.
Figure 4.7:
Removing the Transportation Lock - step 4
10. The threaded bar (1) remains until removing
the lower transportation lock. Afterwards
remove the threaded bar from the top.
Figure 4.8:
Removing the Transportation Lock - step 5
11. Remove three nuts M 8 (2).
12. Remove the threaded bolts M 8 (1).
13. Loosen the twelve screws M 5 x 12 at the
lower transportation lock plate (3). Secure
the lower transportation lock plate against
falling.
14. Remove lower transportation lock plate (3),
the brass nut (4) and lower fixture flange (5).
15. Loosen the threaded bar from the brass nut
(4).
Continue on top of the cryostat:
16. Remove the threaded bar from the top.
WARNING: Heavy Weights
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...