118
ZTKS0156 / Z31820 / Rev.: 03
Maintenance
7.4.3
Replacement of the Rotary Valve and the Cold Head
i
For removing the cold head an applicable step is necessary. This step must be stable,
not too high and without pitfalls.
i
Do not disconnect the cable from the S2 port at the feedthrough plate during the entire
procedure.
Do not switch off the External Temperature Control Unit.
Figure 7.5:
Replacing the Rotary Valve - step 1
1.
Stop the Cryogenic Refrigerator.
The temperature will increase and the
External Temperature Control Unit will give
an audible and a visual alarm.
2.
Stop the audible alarm of the External Tem-
perature Control Unit (refer to the supplied
manual of the External Temperature Control
Unit).
3.
Remove the noise protection cover from the
cold head.
4.
Release the connection line (2) from the
cold head (1).
Figure 7.6:
Replacing the Rotary Valve - step 2
5.
Remove the noise protection cap on the top
of the rotary valve column.
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...