69
ZTKS0156 / Z31820 / Rev.: 03
Assembling
4.3.3.5
Mounting the Helium Level Sensor
Figure 4.70: Mounting the Helium Level Sensor - step 1
1.
Clean the O-ring (1) of the helium fill-in turret
with ethanol.
2.
Grease the O-ring (1) with vacuum grease.
3.
Clean the sealing surface for the O-ring.
4.
Mount the O-ring (1) into the sealing sur-
face.
Figure 4.71: Mounting the Helium Level Sensor - step 2
5.
Insert the helium level sensor (1). Make sure
the helium level sensor inserted into the cor-
rect hole of the siphon.
6.
Connect the connector (2) of the helium
level sensor (1) with the connector (3) of the
helium flow system (4).
7.
Insert the helium level sensor (1) into the
gap of the baffle (5) and insert the assembly
into the helium fill-in turret.
Figure 4.72: Mounting the Helium Level Sensor - step 3
8.
Fix the black nut (1).
9.
Calibrate the helium level sensor in the
TopSpin Service Web. Refer to the supplied
manual of TopSpin.
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...