126
ZTKS0156 / Z31820 / Rev.: 03
Disassembling
.
8.2.4
Removing of the Helium Flow System
1.
Follow the procedure in chapter
”Mounting the Helium Flow System” on page 4-
in reverse order.
2.
Restore the helium flow system to its original box.
3.
Remove the helium level sensor out of the helium fill-in turret.
4.
Restore the helium level sensor to its original box.
8.2.5
Removing the Cryogenic Refrigerator Parts
1.
Switch off the External Temperature Control Unit.
2.
Disconnect the three sensor cables at the feedthrough plate.
3.
Follow the procedure described in chapter
in reverse order.
i
For water cooled compressor option only: a small amount of water from inside the com-
pressor will flow out. Provide equipment to absorb water and to dry the floor.
4.
Follow the procedure described in chapter
”Mounting the Rotary Valve and Flex
in reverse order.
14. The KF 40 connection at the vacuum valve will open,
if the RT vessel of the cryostat is at atmospheric
pressure.
15. Wait until the RT Bore is no longer cold and wet
(approximately 3 - 5 days).
16. Remount the current lead and check the temperature
inside with the PT 100 sensors. If they are at room
temperature remove the current lead, the pumping
unit, the flow device and the vacuum valve.
17. Close the cryostat at the vacuum flange with the
protection cap.
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...