65
ZTKS0156 / Z31820 / Rev.: 03
Assembling
Figure 4.62: Mounting the Rotary Valve and Flex Lines - step 11
23. Put the noise protection cap on the top of
the rotary valve column.
Figure 4.63: Mounting the Rotary Valve and Flex Lines - step 12
24. Attach the insulation tube (1) to the connec-
tion line of the rotary valve.
Respect the orientation: put the female part
towards the rotary valve.
25. If necessary cut away 40 mm on the male
side of the insulation to allow free access to
the Aeroquip® coupling.
Figure 4.64: Mounting the Flex Lines and the Connection Line: Alignment
26. Align the rotary valve column (1)
relatively to the magnet stand and
the cryostat (4) as shown.
The connection line (2) between
rotary valve column (1) and cold
head (3) should not be stressed.
27. Remove the plastic cup nuts from
the spikes of the rotary valve
column.
28. Check the rotary valve column is
levelled horizontally. If necessary
adjust the levelling using the three
spikes of the rotary valve column.
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...