ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
6.2
Problem
6.2.1
During Transportation
6.2.2
During Assembling
Continued on next page
Indicator
Possible reason
Solution
By
Tilt Watch / Shock
Watch activated.
Careless transporta-
tion.
1.
Accept delivery with
reservation.
2.
Remark the extent of
damage in the transport
documents.
3.
Start complaint process.
Approved
Customer
Personnel
Visible damage.
Careless transporta-
tion.
1.
Accept delivery with
reservation.
2.
Remark the extent of
damage in the transport
documents.
3.
Start complaint process.
Approved
Customer
Personnel
Indicator
Possible reason
Solution
By
Ceiling height too
low for assembling
on magnet stand.
Site does not meet
the required condi-
tions.
Choose another site that
meets the required condi-
tions.
Bruker Service
Ceiling height too
low for inserting the
helium level sensor.
Site does not meet
the required condi-
tions.
Insert the helium level sensor
before mounting the magnet
stand.
Bruker Service
Helium bore tube
and radiation shield
are not concentric.
Alignment is not
correct.
Check fixation of the
alignment rods.
Bruker Service
Summary of Contents for RS
Page 1: ...NMR Spectroscopy Service Manual RS ...
Page 4: ...4 ZTKS0156 Z31820 Rev 03 ...
Page 8: ...8 ZTKS0156 Z31820 Rev 03 Table of Contents ...
Page 10: ...10 ZTKS0156 Z31820 Rev 03 Contact ...
Page 14: ...14 ZTKS0156 Z31820 Rev 03 Introduction ...
Page 36: ...36 ZTKS0156 Z31820 Rev 03 Transportation ...
Page 86: ...86 ZTKS0156 Z31820 Rev 03 Operation ...
Page 112: ...112 ZTKS0156 Z31820 Rev 03 Troubleshooting ...
Page 132: ...132 ZTKS0156 Z31820 Rev 03 Disassembling ...
Page 134: ...134 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 136: ...136 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 142: ...142 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 144: ...144 ZTKS0156 Z31820 Rev 03 Appendix ...
Page 150: ...150 ZTKS0156 Z31820 Rev 03 Technical Data RS Service This side is intentionally left blank ...
Page 156: ...156 ZTKS0156 Z31820 Rev 03 ...